GE Healthcare · CTHiSpeed182408
After a hardware reset, the table is raised and lowered through full range but will not reference
Symptom
After a hardware reset, the table is raised and lowered through full range but will not reference. Flashings 8s would not go away in the elevation direction.
Error message
Error: 182408 Velocity tracking error. Probable Problem could be 180V Power is LOW. See Long Description for complete error Error: 182408 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 83 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 11:57:28:685 Log Series: 35 HW/SW error: Velocity tracking error. Actual Velocity 0 (.1 mm/sec) Expected Velocity 400 (.1 mm/sec) Tol 320 (.1 mm/sec) Cur Pos 3276 (.1 mm) Elev Direction: 1 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2372485 ETC IF (Interface or TNC Bd.) Replaces 2229871 See long description for details 2372485 ETC IF (Interface or TNC Bd.) Replaces 2229871 Part #: 2372485 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: ETC INTERFACE W/ AZ60A FRU: Yes Status: Orderable Comments: FLS - ETC INTERFACE W/ AZ60A. ADDED PER PCN 220857. REPLACES P/N 2229871. USE UP EXISTING STK. MS20544 5/1/03. ONE USE LIGHTSPEED PLUS VC:611695-SCI, RAPID CITY, SD CONTROL PCB 46-264368G3
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.