Pular para o conteúdo principal
GE Healthcare · MRSigna2241520

Will Not Scan Due To Gradient Errors

Symptom

Will Not Scan with Gradient Errors.

Error message

2241520 The GIP reported a X Axis Current Distortion Fault Tue May 16 17:26:51 2000 Error: 2241520 Host: MRCV_OW0 Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 2231 The GIP reported a X Axis Current Distortion Fault. Another fault reported in the log could be the cause of this fault. Possible causes: GRAM Tuning, miniGRAM, components of the X Axis GPM, GIP, or cables.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues