GE Healthcare · CTHiSpeed200302181200302313200300013

System State cannot restore the Calibration Database

Symptom

System State cannot restore the Calibration Database See LD for gesys_host.log example System State cannot restore the Calibration Database after a software load with error sequence Error: 200302181 Error: 200302313 Error: 200300013 Wed Oct 26 15:41:49 2005 Error: 245522 Suite: CTO1 Host: SBC0 Process: dataacq File: blocked_ref_info.m 1.13 Method: blocked_ref_info.m Line: 648 Function: Data Acquisition : SBC Processing Scan: 0/0/0 Type: Not scanning Scan Seq: 0 Open to read blocked reference information file failed. Reason: Unix system error : 2 --- No such file or directory. Wed Oct 26 15:41:49 2005 Error: 245557 Suite: CTO1 Host: SBC0 Process: dataacq File: dataacq_caldb.m 1.13 Method: get_afterglow_group() Line: 199 Function: Data Acquisition : SBC Processing Scan: 0/0/0 Type: Not scanning Scan Seq: 0 Failed get the afterglow group from cal database, therefore afterglow processing will be turned off. Wed Oct 26 15:43:21 2005 Error: 200302181 Host: ct01 Process: sysStateABB File: ss_osdep.c Line: 83 System call failed. Error: No such file or directory Command: /usr/g/bin/expandcopy /tmp/sslist /data > /dev/null 2>&1 Wed Oct 26 15:43:22 2005 Error: 200302313 Host: ct01 Process: sysStateABB File: ss_gr_saverest.c Line: 323 Failed to restore the ss_caldb file(s). Wed Oct 26 15:43:22 2005 Error: 200300013 Host: ct01 Process: sysstategui File: sysStProc.c Line: 343 The sysStateABB -restoreCal process returned a bad status of 255

Error message

Error: 200302181 System call failed. Error: No such file or directory Wed Oct 26 15:43:21 2005 Error: 200302181 Host: ct01 Process: sysStateABB File: ss_osdep.c Line: 83 System call failed. Error: No such file or directory Command: /usr/g/bin/expandcopy /tmp/sslist /data > /dev/null 2>&1

Error: 200302313 Failed to restore the ss_caldb file(s). Wed Oct 26 15:43:22 2005 Error: 200302313 Host: ct01 Process: sysStateABB File: ss_gr_saverest.c Line: 323 Failed to restore the ss_caldb file(s).

Error: 200300013 The sysStateABB -restoreCal process returned a bad status of 255 See LD for full Error Error: 200300013 The sysStateABB -restoreCal process returned a bad status of 255 Wed Oct 26 15:43:22 2005 Error: 200300013 Host: ct01 Process: sysstategui File: sysStProc.c Line: 343 The sysStateABB -restoreCal process returned a bad status of 255

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.