Pular para o conteúdo principal
GE Healthcare · CTLightSpeedHiSpeed185122182902

LightSpeed 4.X - GRE

Symptom

Cradle will not move in or out of the Gantry. No c-pulse is being detected from longitudal encoder Cradle will not move in or out of the Gantr . No c-pulse is being detected from longitudal encoder y

Error message

Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Power is OFF

Error: 182902 Emergency Off button has been pressed Sun Sep 5 16:04:39 2004 Error: 182902 Suite: RP21 Host: ETC Process: Table Controls File: tc emergency.c 1.6 2/16/99 Method: No Method line: 89 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 4016 Log Series: 22 Emergency Off button has been pressed

Applies to

Models
LightSpeed, HiSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2216522 CHANNEL EXTRUSION. See the LD for part comments Part #: 2216522 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: CHANNEL EXTRUSION FOR TAPE SWITCH FRU: Yes Status: Orderable CHANNEL EXTRUSION FOR TAPE SWITCH(46-297698P3). FRU = 2 THIS PART MUST ALWAYS BE ORDERED WHEN PART# 46-297698P3 IS ORDERED . 46-297698P3 27" HIGH PROFILE Tape switch See the LD for tape switch jpg Part #: 46-297698P3 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 27" HIGH PROFILE BEAD, RIBBON SWITCH FRU: Yes Status: Orderable TABLE TOP COVER TAPE SWITCH. FRU = 2. *P3 WILL REPLACE THE P2 WHEN P2 STOCK IS DEPLETED. FE: IF YOU ARE REPLACING 46-297698P2 YOU MUST ORDER BOTH 2216522 AND 46-297698P3
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.