Pular para o conteúdo principal
GE Healthcare · MRExcite2225536

Cannot Scan With TR and Dynamic Disable Faults.

Symptom

Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.

Error message

Spurious DD/TR Driver Fault

2225536 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. See Long Description Error: 2225536 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID PMH.m 1.29 Method: spi_tr_mon Line: 6220 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. It cleared itself before the specific problem could be read. Possible causes: noise on the dynamic disable board or a bad board.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
5118952 1.5T Ground Isolation Filter See Long Description Description in FEMC: Ground Isolation Filter These are the Isolation Boxes that connect to the BODY QUADRATURE HYBRID ASSEMBLY
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues