Pular para o conteúdo principal
GE Healthcare · MRSignaExcite

26567 Disk Space Problems

Symptom

Disk Space Problems See long description. LX systems will sometimes generate error messages indicating disk space problems. The message will recommend a system shutdown and reboot, which will run storelog to clear the problem. On occasion this will not work, and disk space limitations will still exist. There is a helpful file generated in these situations which can lead to the correct fix action. Check the /usr/g/service/log/diskfull.log file for LX disk space problems. LX systems will sometimes generate error messages indicating disk space problems. The message will recommend a system shutdown and reboot, which will run storelog to clear the problem. On occasion this will not work, and disk space limitations will still exist. There is a helpful file generated in these situations which can lead to the correct fix action. Check the /usr/g/service/log/diskfull.log file for LX disk space problems. {sdc@b3a}[1] cd /usr/g/service/log <enter> {sdc@b3a}[2] more diskfull.log <enter>

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues