GE Healthcare · CTLightSpeed1831006500118310833003

Scanning aborted with unexpected Loss Of High Voltage.

Symptom

Scanning aborted with unexpected Loss Of High Voltage.

Error message

Error: 183100 Two consecutive kV samples out 10.0 % tolerance Open Long Description to see error example Tue Jun 7 03:35:46 2005 Error: 183100 Suite: GEC2 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 521 Function: XRAY GENERATION : KV Loop Scan: 65001/11/14 Type: Axial Exception Level: Secondary Time: 03:35:45:575 Log Series: 223 Two consecutive kV samples out 10.0 % tolerance. Selected technic: 140 kV, 56 mA Sample1 = 93.919 kV. Sample2 = 0.000 kV.

Error: 183108 KV fault detected: Cathode shoot-thru. See long description for full error Sun Jun 5 23:17:05 2005 Error: 183108 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 248 Function: XRAY GENERATION : KV Loop Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:05:545 Log Series: 424 KV fault detected: Cathode shoot-thru. Fault latch address = ffcff3 hex, Bit = D3

Applies to

Models
LightSpeed
Product
HSA RP 1.X HSA RP 1.X with Zeus HSA Gantry and Sun Genesis Console FMI 25246 HSA/RP 1.6 Y2K Upgrade Special Tools Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2122960 COMMON INVERTER HSA / RP-EMC FOR ANODE OR CATHODE (CTi Inverter) 2122960 COMMON INVERTER HSA / RP-EMC FOR ANODE OR CATHODE (CTi Invertor) 2122960 COMMON INVERTER HSA / RP-EMC FOR ANODE OR CATHODE (CTi Inverter) Part #: 2122960 Part Type: GP REPAIRABLES Part Weight: 30.86 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: COMMON INVERTER FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution