Pular para o conteúdo principal
GE Healthcare · CT2300000675000416800015000230000114230000713

Lightspeed 1.X - Octane

Symptom

Scanner Hardware Stopped Scan with DIP and view length errors. Scanner Hardware Stopped Scan with DIP and view length errors

Error message

Error: 230000067 Unexpected event occurred during scan data transfer See the long description for the full error code Sat Apr 17 20:34:13 2004 Error: 230000067 Host: darc Process: dataacq File: dip_thread.cxx Line: 390 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Xray-on Static Scan: 50004/4/1 Scan Seq Id: 168000 Unexpected event occurred during scan data transfer. Event bit: 7 Count: 1 Description: Global bus error: read attempted from the Test Data FIFO Scan aborted. OR Tue Mar 6 07:13:50 2007 Error: 230000067 Host: darc Process: dataacq File: vdip.cxx Line: 817 Function: Data Acquisition : Scan Data Acquisition Scan Type: Cine Scan: 50004/6/1 Scan Seq Id: 15000 Unexpected event occurred during scan data transfer. Event bit: 22 Count: 1 Description: DMA interrupt was not cleared Scan aborted.

Error: 230000114 DIP scan relay test: Initial clearing of DIP interrupt status register failed See the long description for the full error code Mon Aug 1 00:50:35 2005 Error: 230000114 Host: qxi1_sbc Process: dataacq File: dip_if.cxx Line: 1850 Function: Data Acquisition : SBC Processing Scan Type: Not scanning Scan: 0/0/0 Scan 

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2160800 DAS INTERFACE PROCESSOR DIP Card for O2 cpu See LD for more info 2160800 DAS INTERFACE PROCESSOR DIP Card for O2 computer Part #: 2160800 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: DAS INTERFACE PROCESSOR FRU: Yes Status: Orderable Comments: FLS - DAS INTERFACE PROCESSOR USED ON HELIOS. MS08992 9/3/98 BEING REPLACED BY NEW VERSION 1ST QTR 2001.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues