Pular para o conteúdo principal
GE Healthcare · MRExcite22480101913212248102101512248239

PSD download fails. 3T Excite G3.

Symptom

Protocol PSD Download Fails. Protocol PSD Download Fails

Error message

2248010 A/D Offsets is out of Valid Range See Long Description 2248010 File: DRF Line: 350 Diagnostic: RRF Calib Iteration: 1 of 1 Action: 1 Task ID: 0xa55688 Time: 191321 Object: APS A/D Offsets is out of valid range The calculated value is 1, valid range is 127,-128 (rcvr=0) Suspect analog receiver problem, please run Diagnostics!

2248102 Received invalid cm packet 10151, scan state = 6 Software Error. Error: 2248102 Host: amr12 Process: diagCoord File: processPkt Line: 896 Diagnostic: RRF Calib Iteration: 1 of 1 Action: 1 Task ID: 0xc2a4f8 Time: 191321 Object: APS Received invalid cm packet 10151, scan state = 6 Software error, please perform a TPS Reset.

2248239 ADC Offsets Calibration Validation Failed. The Acquired ADC Offsets are out of Tolerance for the Following Receivers. ADC offsets calibration validation failed. The acquired ADC offsets are out of tolerance for the following receivers. Rx=2 Tol = +/-128 A TPS reset may be necessary.->

Applies to

Models
Excite
Product
3T Excite G3 Software ONLY
Software
G3 For 3T Excite only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues