GE Healthcare · CTLightSpeed182407
Table will not raise
Symptom
Table will not move up, especially when in full down position. See long description Table will not raise or move up, especially when it's in the full down position and this could be intermittent along with noises noticed by the customer.
Error message
Error: 182407 Velocity error detected when trying to find the Index Pulse Sun Aug 29 13:25:59 2004 Error: 182407 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 174 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 13:25:59:690 Log Series: 34 HW/SW error: Velocity error detected when trying to find the Index Pulse. Hardware Inhibit Switches: 0xc0d8 Elev Direction: 0 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-288170G1 ELEV/TILT AMP See LD for more 46-288170G1 ELEV/TILT AMP Part #: 46-288170G1 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEV/TILT AMP FRU: Yes Status: Orderable Comments: FLS - B: THE PROBLEM IS USUALLY WITHTHE 24 V RELAY. PLEASE HAVE THE FE TRY PART 2120791-9. MSA6521 5/29/01 DELISTED T3202CF 12/00 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD
- Area
- CT
Full cause & step-by-step resolution