GE Healthcare · CTHiSpeed186357
Difficulty in positioning the cradle.
Symptom
Difficulty in positioning the cradle. Cradle will not drive normally and scan aborts with errors.
Error message
Error: 186357 Cradle slip was detected in response to CRADLE CMD Code: 12-1003-03 See the long description for the full error code Error: 186357 Cradle slip was detected in response to CRADLE CMD Cradle Error. Code: 12-1003-03 Error detected by: TGP (TP) Recommendation: ERRORS DETECTED BY TGP BOARD (FOR SYSTEM VERSION 5.5 OR LATER) Cradle Req Error: 3 Code: 12-1003-03 Name: Cradle Slip. Class: Error Description: Cradle slip was detected in response to CRADLE CMD. Potential Cause / FRU: 1.Patient movement during cradle movement or some obstacles on bottom of cradle or cradle roller. 2.Cradle Encoder
Applies to
- Models
- HiSpeed
- Product
- HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
- Component
- 2209832-2 table encoder assembly - used on YMS Xi (replaces 2209832) See Long Description for Details 2209832-2 replaces 2209832 table encoder assembly Part #: 2209832-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ENCODER ASSY FRU: Yes Status: Orderable Part #: 2209832 Part Type: GP CONSUMABLES Part Weight: 0.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9235SL ENCODER ASSY FRU: Yes Status: Orderable Comments: FLS - ENCODER ASM U/O CT HISPEED-YMS LXI,FXI,DXIM3 FW20 '98 MS1101 3/98 TABLE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required