GE Healthcare · CTLightSpeed185557
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 185557- hemrc serial link is down. Hemrc_+12V failure detected. Possible causes: HEMRC fuse (F5), Interconnects, HEMRC AC Drive, HEMRC Control Board See Long Description for complete error Mon Jul 24 08:41:32 2006 Error: 185557 Host: OBC Process: Generator File: scanbusDriver.c 1.5 Line: 243 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:28:915 Log Series: 31 Could not send HEMRC message in the allotted time. HEMRC serial link is down. HEMRC_ISO_+12V Failure detected. Possible causes: HEMRC fuse (F5), Interconnects, HEMRC AC Drive,
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
- Area
- CT
Full cause & step-by-step resolution