GE Healthcare · MRExcite2248602224860622238172249410236392249412

Unable to Download or Reset The TPS MGD or ISE.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2248602 PDU in Stop Mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2248602 PDU is in Stop mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed.

2248606 The air temperature in the System Cabinet is greater than 40'C (104F). 2248606 GOC-3TMR NSP SCP:IrfPciHWO 1275 The air temperature in the System Cabinet is greater than 40'C (104F). If the air temperature exceeds 50'C (122F), scanning will be inhibited. Check equipment room temp, if warmer than normal contact site maintenance, else call GE Service.

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2249410 A PCI error occured. The PCI Parity Error PERR was asserted. Error: 2249410 File: excLog.c Line: 1132 Task ID: 0x3ac1d68 Time: 23639 Object: AGP A PCI error occured. The PCI Parity Error PERR was asserted. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x88 Raven Err Address Reg: 0x0; Raven Err Attribute Reg: 0x0. Additional information in /usr/g/service/log/vxPciLg-5c2c

2249412 A PCI error occured. The PCI Parity Error PERR was asserted. 2249412 

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2358122-2 IRF2 Board
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.