Pular para o conteúdo principal
GE Healthcare · MR225470222259352254703

Gating does not work and message returned on system boot or TPS reset

Symptom

Gating does not work and the message The PAC does not respond displayed during system boot or TPS reset. This problem was found while installing this new system. Ran Diagnostics --> Peripherals --> PAC Diagnostics --> PAC Power up Test and SPU-->PAC Loop back Test, both of which failed with error codes 2254702 & 2254703. Measured PAC power at rear of SSM module and found it to be okay. This problem was found while installing this new system. Ran Diagnostics --> Peripherals --> PAC Diagnostics --> PAC Powerup Test and SPU-->PAC Loopback Test, both of which failed with error codes 2254702 & 2254703. Measured PAC power at rear of SSM module and found it to be okay.

Error message

2254702 PAC DIAGNOSTIC TIMEOUT. The PAC did not return a result to the SPU for the last PAC diagnostic test in the allotted 8 second time. Try running PAC datapath tests or other PAC diagnostics. Extended Error: Verify that the PAC module has power. Run SPU-->STIF and SPU-->PAC loopback tests to see if the MGD chassis can talk to the PAC module. Check fiber optic connections between the STIF and the PAC. Possible cause(s): PAC

2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. 2225935 The PAC does not respond. Check PAC power and fiber optic connections, run PAC diags. This error condition adversely affects gated scans only. 2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. This error condition adversely affects gated scans only.

2254703 SPU TO PAC LOOPBACK FAILED. The SPU did not receive data back from the PAC during the SPU to PAC loopback test. Incorrect count of data received 0. Data count expected 8 --> Extended Error: The SPU did not receive data back from the PAC during the SPU to PAC loopback test. Incorrect count of data received. %d. Data count expected %d. Possible cause(s): PAC

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR5-HFO5 Open Speed with Excite Product
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues