The ConnectPro Work Llist comes up empty when updated on the MR system
Symptom
The ConnectPro Work List comes up empty when updated on the MR system see long description LX system running 9.1 M4 software. Just connected up to an INTERWORKS hisris Broker. The ConnectPro worklist comes up empty when queried from the MR. lx-mr 91# su - sdc {sdc@lx-mr}[1] getver Hostname : lx-mr Build number for MrpApps is 91.29I_M4_0311.b Build number for PostSdC is 91.29I_M4_0311.b Build number for cclass is 91.29I_M4_0311.b Build number for driverSupport is 91.29I_M4_0311.b Build number for insite is 91.29I_M4_0311.b Build number for install is 91.29I_M4_0311.b Build number for os_cd is 6_15.4.a Build number for patch_91_M4_02 is 91.29I_M4_0311.b.PB Build number for patch_91_M4_03 is 91.29I_M4_0311.b.PB Build number for patch_91_M4_05 is 91.29I_M4_0311.b.PJ Build number for patch_91_M4_07 is 91.29I_M4_0311.b.PG Build number for patch_91_M4_09 is 91.29I_M4_0311.b.PC Build number for patch_91_M4_11_0Connectivity is 91.29I_M4_0311.b.PK Build number for patch_91_M4_12 is 91.29I_M4_0311.b.PG Build number for patch_91_M4_14 is 91.29I_M4_0311.b.PJ Build number for patch_91_M4_15 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_16 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_17 is 91.29I_M4_0311.b.PL Build number for patch_91_M4_18 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_19 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_20 is 91.29I_M4_0311.b.PK Build number for pc is 91.29I_M4_0311.b MR Software release: 9.1.0311b
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.