Pular para o conteúdo principal
GE Healthcare · MRExcite2248662

New Install Getting MDS Failures After Booting.

Symptom

New Install Getting MDS Failures After Booting. New install, when system was coming up it would indicate an error of SCP auto config MDS failure

Error message

2248662 MDS link failure. The SCP failed to receive an MDS transmission echo. Tue May 25 10:38:44 2004 2248662 1 MROP NSP SCP:MdsAutoConfig 1443 MDS link failure. The SCP failed to receive an MDS transmission echo. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0000 Transmitted byte: 0x10

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues