Coil id Light is Always Green

Symptom

Coil ID Light is Always Green See Long Description Coil ID light is always green. GE Healthcare SMILe's and Service Notes are copyright (c) to General Electric Company Inc with all rights reserved. SMILe's and Service Notes issued by GE Healthcare are proprietary to GE Healthcare. Disclosure to persons not employed by GE Healthcare or not in possession of a Advanced Service Agreement with GE Healthcare is prohibited. SMILe 8/23/05 Subject: Coil ID failures with 8 Channel Switch. Effectivity: All Signa Excite 2 (11.x) systems equipped with 8 Channel Switch (part number 2362995-2). Problem Description: The 8 Channel Switch can fail so that the LPCA green Coil ID LED is illuminated at all times, even when a coil equipped with Coil ID is not connected. Oftentimes to recover it is necessary to cycle power to the Driver Module and perform a Reset TPS to allow scanning.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2362995-2 Eight 8 Channel Switch
Area
MR