Unable to Scan after Installing the Jumper into J2 of the PDU Control Board which Allowed the PDU To Power On

Symptom

Unable to Scan after Installing the Jumper into J2 of the PDU Control Board which Allowed the PDU To Power On Cannot turn on Gradients or RF amp. PDU had 480vac incoming power but EMO reset button was not active. In the back of the PDU is a board called PCBA that needs 12vdc in order to activate some relays, which allow the EMO button to be active.

Error message

2248602 PDU in Stop Mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2248602 PDU is in Stop mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed.

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.