Pular para o conteúdo principal
GE Healthcare · MRExcite2262605107521

PHPS CAN Diagnostic Fails

Symptom

PHPS CAN Diagnostic Failure.

Error message

2262605 A Failure was encountered uploading parameter 0x20 from the PHPS Motor Amplifier. Please try again or try resetting the Motor Amplifier through the Diagnostic Screen. Failure code is 26.-> 2262605 ghmr-ow0 diagCoord getPhpsVal 445 Diagnostic: PHPS CAN Iteration: 1 of 1 Action: 1 Task ID: 0x3d5c6c8 Time: 107521 Object: SCP A Failure was encountered uploading parameter 0x20 from the PHPS Motor Amplifier. Please try again or try resetting the Motor Amplifier through the Diagnostic Screen. Failure code is 26.->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
5109368 Patient Handling Power Supply Module (PHPS). 5109368 Patient Handling Power Supply Module (PHPS). For Laser alignment light systems. 5109368 Patient Handling Power Supply Module PHPS
Area
MR