Pular para o conteúdo principal
GE Healthcare · MRExciteSigna2248217284717422473392848379

RRF Calibration Error in RRF Data Path Diagnostics

Symptom

RRF Calibration Error in RRF Data Path Diagnostics

Error message

2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10

2247339 Receiver Calibration Diagnostics Failed Fri Jul 18 05:59:48 2003 2247339 2 b2a diagCoord rrfdpDiags.c 507 Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1 Task ID: 0x3a8ed38 Time: 2848379 Object: APS Test Task Receiver calibration diagnostics failed. Please perform a TPS reset.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2366218-2 UTNS3 See Long Description 2366218-2 UTNS3 EMC ED2 Compliant .
Area
MR

Full cause & step-by-step resolution