The Erbtec Amp Will Not Go To Operate Mode Due To Processor Board Failure.
Symptom
Unable to scan with Erbtec errors
Error message
ERBTEC AMPLIFIER Fault 06 Stepping motor coil is apparently open ERBTEC AMPLIFIER Fault 06. Stepping motor coil is apparently open; very possibly an open in the Motor Cavity Wiring Harness or connector; less possible cause is a component failure on the Processor Board
Applies to
- Models
- Signa
- Product
- Signa Advantage Includes 4.x and 5.x systems. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 5.X
- Component
- 46-305000P4 Erbtec Amplifier PROCESSOR BOARD For 1.0/1.5T Amplifier NOTE 1. CAREFULLY REMOVE U16 CHIP (68HC11) AND U17 CHIP (EPROM) FROM ITS SOCKETS BY GENTLY PRYING UP BOTH ENDS SIMULTANEOUSLY (USE AN IC REMOVING TOOL IF POSSIBLE). PROTECT AND SET ASIDE FOR REPLACEMENT IN NEW PROCESSOR BOARD. CAREFULLY REMOVE U16 CHIP (68HC11) AND U17 CHIP (EPROM) FROM ITS SOCKETS BY GENTLY PRYING UP BOTH ENDS SIMULTANEOUSLY (USE AN IC REMOVING TOOL IF POSSIBLE). PROTECT AND SET ASIDE FOR REPLACEMENT IN NEW PROCESSOR BOARD.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.