Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230000067500041680001500023000006450003

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 230000067 Unexpected event occurred during scan data transfer See the long description for the full error code Sat Apr 17 20:34:13 2004 Error: 230000067 Host: darc Process: dataacq File: dip_thread.cxx Line: 390 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Xray-on Static Scan: 50004/4/1 Scan Seq Id: 168000 Unexpected event occurred during scan data transfer. Event bit: 7 Count: 1 Description: Global bus error: read attempted from the Test Data FIFO Scan aborted. OR Tue Mar 6 07:13:50 2007 Error: 230000067 Host: darc Process: dataacq File: vdip.cxx Line: 817 Function: Data Acquisition : Scan Data Acquisition Scan Type: Cine Scan: 50004/6/1 Scan Seq Id: 15000 Unexpected event occurred during scan data transfer. Event bit: 22 Count: 1 Description: DMA interrupt was not cleared Scan aborted.

Error: 230000064 TAXI data violation detected by forward error correction See the long description for the full error code Sat Apr 17 20:26:43 2004 Error: 230000064 Host: darc Process: dataacq File: dip_thread.cxx Line: 411 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2205268 15V 1.7A power supply - Used In 1.X, 2.X, 3.X, 4.X, and 5.X gantrys 2205268 15V 1.7A power supply Used 1.X, 2.X, 3.X, 4.X, 5.X gantrys Part #: 2205268 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: POWER SUPPLY 15V DC 1.7 AMPS FRU: Yes Status: Orderable Comments: FLS - POWER SUPPLY 15V DC 1.7 AMPS U/O HELIOS GANTRY MS08992 10/5/98 USED ON HPOWER PROGRAM, PER RSPL. MS20544 ON 2/12/03. PART TRANSFERRED TO SSP 04/21/00 MS24085 PER KINH LEE CT MFG. THIS POWER SUPPLY IS FOR BOTH THE TRANSMITTER, (MOUNTED TO SLIPRING), AND ALSO RF RECEIVER (MOUNTED TO STATIONARY GANTRY). MS10604 001218 NOTE: F/O CABLE IS 2261709
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues