Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed2900090021183997163967569

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

System fails to reconstruct images. System fails to reconstruct images

Error message

Error: 290009002 Error failed in opening file See long description for full error Mon Jul 9 17:34:45 2007 Error: 290009002 Host: darc Process: recon_control File: IsrUpdateCache.cxx Line: 153 Error failed in opening file. file name = /usr/g/queue/priority3_jobs/FCCC.3776.2.1177.1183997163.967569.pro_cine error msg = No such file or directory

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS See LD for more info 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS Part #: 5114772 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DARC Computer - Dual Disk H16 MDAS, GDAS FRU: Yes Comments: Added per GOC4 RSPL Replace the DARC2: 1) If you get a Westville DARC P/N 5114772 load the DARC software to recover. 2) If you get Jarrell p/n 5147442 perform LFC using 06MW29.7 software.
Area
CT

Full cause & step-by-step resolution