GE Healthcare · CTLightSpeed200352903200002372290009007
LightSpeed 5.X (60kW)
Symptom
Recon self test failure on boot up. Recon self test failure on boot up
Error message
Error: 200352903 Reconstruction errors detected See Long Description for error. Fri Jul 29 20:27:00 2005 Error: 200352903 Host: whct Process: reconSelftest File: reconSelftest.c Line: 200 Reconstruction errors detected Had Recon Retry on image 414/4/1, the recon job did not succeed errors in 1 images during selftest --------------------------------------------------------------------------- Fri Nov 17 12:32:05 2006 Error: 200352903 Host: ct04 Process: reconSelftest File: reconSelftest.c Line: 201 Reconstruction errors detected Had [IG 1] image checksum mismatch errors in 4 images during selftest Error: 200002372 The process reconSelftest took too long to start up. See Long Description for complete error. Mon Nov 5 15:42:38 2007 Error: 200002372 Host: vcsvct Process: cupMonitor File: ssProcStart.c Line: 856 The process reconSelftest took too long to start up. Error: 290009007 Timeout waiting for image available. Number of retry so far: 1. IG node: 0. Sun May 14 14:05:42 2006 Error: 290009007 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 5208 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_HELICAL_RECON Scan: 378/2/2 Image: 378/2/39 Timeout
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2362872 Westville IG Module used on GOC3 and GOC4 consoles. See Long Description for details. 2362872 Westville IG Module used on GOC3 and GOC4 consoles. Part #: 2362872 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 19.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: IMAGE GENERATION COMPUTER - GRE FRU: Yes Status: Orderable Comments: **FE Assist** applies to CT Not Pet systems. If no stock is available order a Jarrell IG 5147443 to replace a Westville IG (2362872), you MUST order p/n 5193989 software release 06MW29.7, and perform a LFC. See Service note SN06101010A. PET must receive the Westville IG.(12/12/06 212001261) *** 2ND OPINION NOTES *** See the following URL for additional command line diagnostic troubleshooting: http://gemsvst4p.med.ge.com:7005/facets/CT/GRE%20command%20line%20diag%20test.pdf?
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.