2274021 Coil Fault Short Circuit A multicoil bias fault was detected due to Coil cable.

Symptom

Cannot scan with HNS coil with Error 2274021 Coil Fault - Short Circuit A multicoil bias fault was detected See Long Description Cannot scan with HNS coil with the error shown below logged. Mon Mar 1 09:22:21 2010 Error: 2274021 Host: WBH_RAD_MR11 Process: NSP File: SCP:RfHubCanHWO::RfBias Line: 4080 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HNS CTL123 Scan: 4637/1 Coil Fault - Short Circuit A multicoil bias fault was detected. Connector: Port 2 (P2) Channels faulted: 0x100 First channel: 9 of 32, counting from 1 Fault value: -4777 mV, Channel: 9-> Mon Mar 1 09:28:02 2010 Error: 2274021 Host: WBH_RAD_MR11 Process: NSP File: SCP:RfHubCanHWO::RfBias Line: 4080 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HNS- CTL12 Scan: 4637/1 Coil Fault - Short Circuit A multicoil bias fault was detected. Connector: Port 1 (P1) Channels faulted: 0x100 First channel: 9 of 32, counting from 1 Fault value: -4840 mV, Channel: 9->

Error message

2274021 Coil Fault - Short Circuit A multicoil bias fault was detected See Long Description Error: 2274021 Host: WBH_RAD_MR11 Process: NSP File: SCP:RfHubCanHWO::RfBias Line: 4080 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HNS CTL123 Scan: 4637/1 Coil Fault - Short Circuit A multicoil bias fault was detected. Connector: Port 2 (P2) Channels faulted: 0x100 First channel: 9 of 32, counting from 1 Fault value: -4777 mV, Channel: 9

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
2424269 16 CH HNS DVMR SYS CABLE FRU 2424269 16 CH HNS DVMR SYS CABLE FRU
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.