GE Healthcare · MROptima2261715222755622569252223817
Fails TPS Reset when booting up to Applications or when Resetting TPS due to open fuse on DI board in HEC Signal Box.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2261715 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached an alarm level. See Long Description Error: 2261715 Host: lx-mr Process: NSP File: SCP:CoolingHWO Line: 2901 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: FSE-XL Coil: HDCTL 456 Scan: 521/9 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached an alarm level. Scanning will be inhibited until coolant serviced. Please contact service.--> Extended Error: Order coolant deionization kit and schedule service. If the resistivity is 0 kOhm-cm, check the wiring between the sensor in the gradient coolant reservoir and the signal box. 2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. *** 2256925 StartupCondition error on SCP resulted in TPS reset failure. See Long Description 2256925 MOI_MR LxMxConsumer services/initialization/ServerInitManager 639 Server Name: SC StartupCondition error on SCP resulted in TPS reset failure. 2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisR
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5266827 Heat Exchanger Fuse Kit 5266827 Heat Exchanger Fuse Kit
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.