Port A on the LPCA has a partial short between pins 2 and 3 of the K column.

Symptom

Any receive only coil connected to Port A immediately generates error 2258107 Coil ID light goes through normal sequence (brief red, then green) then error 2258107 is logged.

Error message

2258107 Host coil validation error Body Transmit Enable mismatch detected Error: 2258107 Host: MR3T Process: coilCK File: coilCK scp.c Line: 1026 Exception Class: Unknown Function: Host: Coil Check Host coil validation error: Body Transmit Enable mismatch detected. Connector: Multicoil A Port Coil: HD Shoulder Array by GE - A/B port

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
5111020-50 HDMR2 A plus Bezel and cable harness.
Area
MR

Full cause & step-by-step resolution