Optical fiber Run 4034 is Loose at J1 and J2 of the OE Conversion box

Symptom

2251053 Gradient framing error The GIP reported a data framing error stopping scanning, all axis reported fault.

Error message

2251053 Hardware Error Gradient data framing error. The GIP reported a data framing error. Check the fiber optic clock cable between the Systems Cabinet and GIP. Check the STIF, SRF, and GIP Boards. Run fiber optic diagnostics.

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution