GE Healthcare · CTLightSpeed260003142500012600031185017250066
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260003142 The Filter failed to move to the requested position See long description for more details Sun Apr 8 06:41:17 2007 Error: 260003142 Host: HSC Process: CCB Firmware File: filter_move.c Line: 104 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50001/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:41:17:580 Log Series: 46 The Filter failed to move to the requested position. Error: 260003118 CAM A drive tracking detected the current motor position is out of the tolerance band for a successfull move See Long Description for complete error. Error: 260003118 CAM A drive tracking detected the current motor position is out of the tolerance band for a successfull move Mon Aug 6 09:44:18 2001 Error: 260003118 Host: HSC Process: CCB Firmware File: ccb_drive_timers.c Line: 560 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50172/4/0 Type: None/Unknown Exception Level: Sec/Soft Time: 09:44:18:540 Log Series: 90 CAM A drive tracking detected the current motor position is out of the tolerance band for a successfull move Start Position = 3995um 165cnt Desired Position = 2um -863cnt Current Moto
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5331736 - Helios Filter Asselmbly FRU (replaces 2291958) Used on Helios 3 and 4 Collimators. See Long Description for more Details Part #: 5331736 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Helios Filter Asselmbly FRU FRU: Yes Status: OrderableComments: Per ECR 2082045, 212008898, 6/16/09. Part #: 2126674 Part Type: GP CONSUMABLES Part Weight: -144548.55 LBR Repairable: No Returnable: No Hazardous Material: N Description: FILTER ASSEMBLY, HELIOS COLLIMATOR FRU: NO Status: Non Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required