GE Healthcare · CTLightSpeed26000745750084
Scanning aborts with beam tracking errors
Symptom
"Scan hardware Stopped Scan" with beam tracking errors. "Scan hardware Stopped Scan" with beam tracking errors
Error message
Error: 260007457 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance See Long Description for complete error Wed Apr 27 18:54:40 2005 Error: 260007457 6 DAS DCB Firmware zaxis.c 564 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50084/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 18:54:40:490 Log Series: 936 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance. Tolerance : 400 umd Control Error : 448 umd Beam Side : B Side
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required