Pular para o conteúdo principal
GE Healthcare · CTLightSpeed11009

LightSpeed 5.X (80/100kW)

Symptom

Site cannot send a specific exam or series to PACS or other dicom hosts. All other exams send without error. See Long Description for more details Site cannot send a specific patient to PACS or other dicom host. All other exams send without error. Error as seen in the GESYSLOG Fri Apr 7 19:00:01 2006 Error: 11046 Host: fchct1 Process: dcserver File: dcspush.c Line: 1381 STORE response received with refused status (0xA700) - Aborting! Fri Apr 7 19:00:01 2006 Error: 11009 Host: fchct1 Process: dcserver File: dcspush.c Line: 1516 Association was aborted with PACS_Live Fri Apr 7 19:00:01 2006 Error: 11080 Host: fchct1 Process: dcserver File: dcscommon.c Line: 242 Connection error: 11283 (PACS_Live) Error as seen in the DCSLOG MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1381) [Dcserver]> STORE response received with refused status (0xA700) - Aborting! MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1516) [Dcserver]> Association was aborted with PACS_Live MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcscommon.c:242) Connection error: 11283 (PACS_Live)

Error message

Error: 11009 Association was aborted with hostname See LD for full details Error: 11009 Association was aborted with hostname Error as seen in GESYSLOG Fri Apr 7 19:00:01 2006 Error: 11009 Host: fchct1 Process: dcserver File: dcspush.c Line: 1516 Association was aborted with PACS_Live Error as seen in the DCSLOG MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1516) [Dcserver]> Association was aborted with PACS_Live

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.