PA Air Too Hot Due to Obstructed Air Flow. 3T RF Amplifier Fails with PA Air Too Hot Fault 47.

Symptom

3T RF Amplifier Fails with PA Air Too Hot Fault 47

Error message

2254261 3T RF Amplifier PA Air Too Hot (fault 47). See Long Description Mon Oct 9 13:20:16 2006 Error: 2254261 Host: MR3T Process: NSP File: SCP:../rf_nb_contro Line: 1491 3T RF Amplifier PA Air Too Hot (fault 47). Possible Causes: obstructed air flow, PSD exceeding safe operating range. Cool 30 minutes, clear air intake and exhaust port, adjust PSD parameters. If the fault persists, replace the RF amplifier. Extended Error : The RF Amp PA Tube experienced high air temperature. Allow the RF Amp to cool for 20 minutes and restart scanning. Make sure room air temperature is <40 C, address with site if necessary. Check air filters and replace if necessary. If fault persists, replace the RF deck.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.