Symptom

Keyboard is not typing the characters correctly See the LD for an example Keyboard is not typing the characters correctly h*elp.my7k*eyboardontwor7k

Error message

Unknown key pressed - found in the /var/log/messages log See long description for details Unknown key pressed -- found in /var/log/messages log Open a unix shell and view the last 500 lines of the HP computer log cd /var/log tail -500 messages | more Nov 24 21:03:48 ct03 kernel: atkbd.c: Unknown key pressed (translated set 2, code 0x98 on isa0060/serio0). Nov 24 21:03:48 ct03 kernel: atkbd.c: Use 'setkeycodes e018 keycode ' to make it known.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
2275756 keyboard - NORTH AMERICAN PC PS/2 See the long description for details Part #: 2275756 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: KEYBOARD, NORTH AMERICAN PC PS/2 FRU: Yes Status: Orderable Comments: FLS - H2 ENGLISH SCIM KEYBOARD. DSM 2262158-9, 2275756, AND 2289697-2 ARE REPLACING 2276441-10. ALL THREE PARTS ARE REQUIRED TO REPLACE 2276441-10. USE EXISTING STOCK FIRST OF 2276441-10 PER PCN 224305. 01/04/02 FOUR PARTS (2262158-9, 2275756, 2289697-2, AND 2289697-8) ARE REQUIRED TO REPLACE ANY OF THE FOLLOWING: 2276441-4, 2276441-5, 2276441-6, 2276441-7, 2276441-11, OR 2276441-12. PER PCN 224305, 224332, AND 224426. 04/05/02
Area
CT

Full cause & step-by-step resolution