GE Healthcare · CTDiscoveryOptima26010723050000
Discovery CT590 / Optima CT580 / RT
Symptom
Axial motor drive fault. Gantry will not rotate
Error message
Error: 260107230 Axial Motor Control (AMC) logic detected GANTRY CONTACT Mon Nov 21 16:49:12 2011 Error: 260107230 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 2142 Mon Nov 21 16:49:12 2011 Host : Table Gantry Processor Ermes # : 260107230 Exception Class : Safe State Severity : Primary File : amc_mgr.c Line# : 2142 Function : AXIAL CONTROL : Axial Control Scan Type : Axial Scan: 50000/1/1 Axial Motor Control (AMC) logic detected GANTRY CONTACT switch FAULT during Gantry Rotating ACCEL command. CALL Service - Emergency Stop of the Axial Drive. (AMC State:5 Cmd:21 Reference:4000 CurVel:0 (.001 sec/rev)).
Applies to
- Models
- Discovery, Optima
- Product
- Discovery CT590 / Optima CT580 / RT Discovery CT590 RT (16 slice CT for Oncology) Optima CT580 RT (8 slice CT for Oncology) Optima CT580 W (16 slice Radiology CT for Bariatric and Interventional, etc.) Freedom RT16 see long description for details Product Name: Discovery and Optima RT Discovery CT590 RT (16 slice CT for Oncology) Optima CT580 RT (8 slice CT for Oncology) Optima CT580 W (16 slice Radiology CT for Bariatric and Interventional, etc.) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo List of Supported Console Types TIO (True-In-One Console) List of softwares the product can run 09BW35.11+SP1.7 10BW06.5+SP1.5 FAQ's GT 1700 Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy GT 650 High Capacity Table 650 lbs. Tube type - D3194T D3193T(green) Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. RT Freedom Service Introduction - SN130902B
- Component
- 2269009 Gantry Contact switch ,2 segment - used on most Lightspeed and above gantries Part #: 2269009 Part Type: GP CONSUMABLES Part Weight: 0.16 LBR Repairable: No Returnable: No Hazardous Material: N Description: GANTRY CONTACT SWITCH,2 SEGMENT FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.