GE Healthcare · CTHiSpeed2600066055000026000680350079
HiSpeed QX/i: Hardware Stopped Scan with Axial Errors
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260006605 Could not send axial drive message in the allotted time. Axial serial link is down. See the LD for the full error code Error: 260006605 Could not send axial drive message in the allotted time. Tue Nov 16 14:20:36 2004 Error: 260006605 Host: STC Process: Axial File: scanbusDriver.c 1.5 Line: 523 Function: AXIAL CONTROL : axial communications Scan: 50000/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:20:37:325 Log Series: 70 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices, .. Error: 260006803 Axial Motor Control (AMC) logic could NOT access the AXIAL drive via the CAN network. See long description for complete error Error: 260006803 Host: STC Process: Axial File: amc_exec.c Line: 3077 Function: AXIAL CONTROL : axial communications Scan: 50079/4/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 05:23:06:435 Log Series: 505 Axial Motor Control (AMC) logic could NOT access the AXIAL drive via the CAN network.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2233402 Axial Drive Motor Controller See Long Description for details Part #: 2233402 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 8.64 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: AX DRIVE MTR CONTRLLER W/CORERS ASSY FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: MUST ORDER FROM HOLT ELECTRONICS. MS26561 DSM 1/7/01 USED ON HPOWER PROGRAM, PER RSPL. MS20544 ON 2/12/03. DO NOT DISCOUNT - PER SHAKUN*** 2ND OPINION NOTES *** Check fuses/connections associated with axial drive circuit; run axial functional diagnostic test and check for failures. If used on HP60
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.