Symptom
Unable to scan with a message on screen that says: Collimator/DAS reset failed.
Error message
Error: 260001085 Communication error: Message not sent to the DCB. Communications with the DCB are down. CAN message could not be sent. Message code: 11 See Long Description for complete error Sun Sep 5 08:05:00 2004 Error: 260001085 Host: OBC Process: HOP File: hop_config_track.c Line: 164 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:04:45:705 Log Series: 30 Communication error: Message not sent to the DCB. Communications with the DCB are down. CAN message could not be sent. Message code: 11 Error: 244584 Collimator/DAS reset failed See Long Description for complete error Error: 244584 Host: qx-ct Process: fwmgr File: HwManager.c Line: 1536 Collimator/DAS reset failed.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2106993-5 FUSE 20 AMPS 500 VOLTS .3 SECONDS See Long Description for details Part #: 2106993-5 Part Type: GP CONSUMABLES Part Weight: 0.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 20 AMPS 500 VOLTS .3 SECONDS FRU: Yes Status: Orderable Comments: FLS - 20AMPS 500V .3SECOND FUSE U/O COST REDUCED GENERATORS SCPU 65/80 MS12746 PER RSPL 5-97 GELS - MAJ LEAD-TIME LE 99/07/26 PA,FUSE 20A 500V .3 SECONDS
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.