Pular para o conteúdo principal
GE Healthcare · MRSigna22238592223852

Can Not Boot Application Software With VRE Errors due to ICN.

Symptom

Can Not Boot Application Software With VRE Errors See Long Description 2249418 Host: LJP MR 01 Process: diagCoord File: mgdArp.c Line: 154 Task ID: 0x7effdf0 Time: 1689 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A Network problem detected. Check MGD subnet network configuration. arpResolve failed for IP 10.0.0.102, numTries 2, delay 100 ticks. status = -1. Or, 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Or; 2223860 VRE reset failed on ICN icn1 with MAC address XX:XX:XX:XX:XX:XX Or, 2223885 Host: hde-mr Process: logermes File: vreBmcUpCheck.pl Line: 251 Exception Class: Shutdown Function: Host: NSP System startup failed due to failure to communicate with the VRE!-> Extended Error: The system startup failure is caused by the failure to communicate with all configured ICNs. Please check that power is applied to all ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.

Error message

2223859 The TPS Reset Failed because of a VRE Reset Failure See long description. 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Extended Error: This indicates that a failure to reset the VRE properly caused the TPS Reset to fail. Please refer to the VRE Troubleshooting Guide for further troubleshooting this issue, which can be found in the MR Service Documentation.

2223852 The APS board reported a communication manager (CM) initialization error. See Long Description Please check the error log for a more detailed error message from the board. Extended Error: MCP820 boards can be identified by the word MCP820 printed on the board extraction lever

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5127452-3 Sun Image Compute Node - Gen 2 without VRF 5127452-3 Image Compute Node - Gen 2 without VRF
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.