Pular para o conteúdo principal
GE Healthcare · MRDiscovery22678352256925

gate-drive over-current Errors Due to XGA Amplifier.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2267835 A gate-drive over-current has occurred on % axis XGA Aug 18 Tue 05:04:17 2267835 NSP SCP:FspMac.cpp 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 A gate-drive over-current has occurred on Z axis XGA. Extended error log: 2267835ERROR: EM_XGD_AS_GD_OC A gate-drive over-current has occurred on %b XGA. If this message persists after TPS resets, the corresponding XGA should be replaced.

2256925 StartupCondition error on XGD StartupCondition resulted in TPS reset failure See Long Description 2256925 SLMAINMR LxMxConsumer services/initialization/ServerInitManager 767 Server Name: SC StartupCondition error on XGD StartupCondition resulted in TPS reset failure.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.