VCC measurement on front Panel of power supply in GIP out of Spec.

Symptom

Unable to Scan with Error 2241608 GIP VCC signal is out of range See Long Description Unable to scan due to error code 2241608, see copy below. Error 2241608, "GIP VCC signal is out of range. Check VCC on the GIP front panel. If voltage is correct, replace the GIP board. Else replace the GIP power supply. Low: 4.70V High: 5.250V Value: 4.744V Axis X" FE read the "Repair Operations Service TIP" shown below and measured the 5 volt test point as being 4.61 volts on the GIP front panel. He attempted to adjust it up but couldn't get it into spec via the voltage adjustment "Repair Operations Service TIP" Title: T3503TR MDS Link Failures (lists basic observation checks, and THEN check front panel voltages...) Test point readings: -VCC: 5.15V (MUST ALWAYS BE SET @5.15V. Can be adjusted on PWR Supply @V1 ADJ) o +12V: 12.00 +/-0.10V o -12V: -12.00V +/-0.10V o +15V: 15.00 +/-0.10V Failure to see the corresponding voltages will more than likely cause the errors mentioned below. MDS Link Failure GIP ...V out of range GIP ...V Supply is out of range

Error message

2241608 GIP VCC signal is out of range. Error 2241608, "GIP VCC signal is out of range. Check VCC on the GIP front panel. If voltage is correct, replace the GIP board. Else replace the GIP power supply. Low: 4.70V High: 5.250V Value: 4.744V Axis N ........"

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2169941 GIP ASSEMBLY FOR SGD 2169941GIP ASSEMBLY FOR SGD
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.