Pular para o conteúdo principal
GE Healthcare · MRExcite22493071081934830200002249418224715611409

Unable to Boot Applications Software with TPS Download Failure due to SCP2 Board in MGD Chassis.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2249307 Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Error seen with MGD Diags: 1081934830 0 1 Wed Apr 14 09:27:10 2004 2249307 2 CARD diagCoord pearlMap.c 1806 Task ID: 0x0 Time: 0 Object: SCP Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Value 0xAAAA not found at address 0xC03010A8 after 20000 msec.->

2249418 Network problem detected. Check MGD subnet network configuration. See Long Description Wed Apr 14 09:27:10 2004 2249418 2 CARD diagCoord mgdArp.c 157 Task ID: 0xf99df8 Time: 981 Object: SCP Network problem detected. Check MGD subnet network configuration. etherAddrResolve failed for network interface motfec0, IP 10.47.1.2, numTries 5, delay 100 ticks. status = -1.

2247156 The IRF2 board was not detected in the MGD chassis Error: 2247156 Host: BOWIE-MRI Process: diagCoord File: Irf.cpp Line: 74 Task ID: 0x3f1fdf8 Time: 11409 Object: APS The IRF2 board was not detected in the MGD chassis. The system will not function without this board.->

2249333 PCI communication error. Could not locate the Shared Memory Anchor. 1081934956 1 1 Wed Apr 14 

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2294300-19 MGD SCPII with CAN PMC and I2C firmware fix 2294300–19 SCP2 BOARD
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues