Application Software does not complete the boot process with TPS Download failure Due to VRE Configuration.
Symptom
Can Not Boot Application Software With VRE Errors See Long Description 2249418 Host: LJP MR 01 Process: diagCoord File: mgdArp.c Line: 154 Task ID: 0x7effdf0 Time: 1689 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A Network problem detected. Check MGD subnet network configuration. arpResolve failed for IP 10.0.0.102, numTries 2, delay 100 ticks. status = -1. Or, 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Or; 2223860 VRE reset failed on ICN icn1 with MAC address XX:XX:XX:XX:XX:XX Or, 2223885 Host: hde-mr Process: logermes File: vreBmcUpCheck.pl Line: 251 Exception Class: Shutdown Function: Host: NSP System startup failed due to failure to communicate with the VRE!-> Extended Error: The system startup failure is caused by the failure to communicate with all configured ICNs. Please check that power is applied to all ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.
Error message
2223859 The TPS Reset Failed because of a VRE Reset Failure See long description. 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Extended Error: This indicates that a failure to reset the VRE properly caused the TPS Reset to fail. Please refer to the VRE Troubleshooting Guide for further troubleshooting this issue, which can be found in the MR Service Documentation.
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.