Symptom

Unable to display any images; exam browser listing in image works is empty and fails to refresh.

Error message

Error: 210000445 Error: saveItem() failed with status X. See Long Description for complete error. Fri Sep 30 14:59:58 2005 Error: 210000445 Host: ct99 Process: dms File: DMFInstall.cxx Line: 236 Error: saveItem() failed with status -1 error message TdbExceptionNotImplem ented- for image path /usr/g/sdc_image_pool/reserved/IMSsecBin16/E613U0_S 2U0_I25U0.1128106798_69694 Thu Jul 30 08:55:26 2009 Error: 210000445 Host: H0140CT1 Process: dms File: DMFInstall.cxx Line: 450 Error: saveItem() failed with status 0 error message DMFCreateSession() failed for image path /usr/g/sdc_image_pool/reserved/IMSse7Q2TVf/E1009U0_S1U0_I1U0.1248958526_266351

Error: 210000426 Unable to install image Exxxxxxxxx with status -1 See Long Description for complete error. Fri Sep 30 14:59:58 2005 Error: 210000426 Host: ct99 Process: dms File: InstallImage.cxx Line: 551 Unable to install image E613S2I25 with status -1

Error: 210000447 unable to connect to session dms_session with error code -2 See long description for details Tue Jan 29 06:51:29 2008 Error: 210000447 Host: gevct2 Process: dms File: DMFInstall.cxx Line: 229 Error: unable to connect to session dms_session with error code -2 error message Sessio

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required