GE Healthcare · CTLightSpeed26000312050042
LightSpeed 5.X (60kW)
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 260003120 An error occurred attempting to home CAM A or B See LD for complete error Mon Jan 10 09:04:41 2005 Error: 260003120 Host: HSC Process: CCB Firmware File: aperture_home.c Line: 193 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50042/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 09:04:41:350 Log Series: 57 An error occurred attempting to home CAM A. An error occurred attempting to home CAM B.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296259P6 Rotating Coupling See Long Description for FEMC and pics 46-296259P6 ROCOM COUPLING_BORES 5 MM pictures Part #: 46-296259P6 Part Type: GP CONSUMABLES Part Weight: 0.10 LBR Repairable: No Returnable: No Hazardous Material: N Description: ROCOM COUPLING_BORES 5 MM, 0.250 IN. FRU: Yes Status: Orderable Comments: FLS - ADAPTER ADDED PER PCN #220365.USED ON GANTRY. PER PCN COMMENTS, FRU WILL BE CHANGED FROM N TO 1. MSA8429 01/15/02 COUPLING OF 2 SHAFTS FOR COLLIMATOR MOTOR MS10147 01/02 FOR ROTATING APTERETURE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.