Pular para o conteúdo principal
GE Healthcare · MRExcite22471532247150224716051176

TPS Reset Fail due to RRF Chassis 5V Digital Power Supply And 5V RF Power Supply.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2247153 RF SUBSYSTEM NOT RESPONDING 2247153 MR1 diagCoord Irf.cpp 236 Task ID: 0x3f1fdf8 Time: 823 Object: APS RF SUBSYSTEM NOT RESPONDING The RRF chassis is not communicating to the MGD chassis. Try TPS Reset. If this problem repeats, call service. Scanning will not be possible. Verify that the RRF chassis is powered on

2247150 The IRF driver function Exciter::ex_set_2nd_rrfdif_clk_en timed out when waiting for the RRF chassis to respond to a request. See "Long Description" 2247150 diagCoord Exciter.cpp 1398 Task ID: 0x3f1fdf8 Time: 868 Object: APS The IRF driver function Exciter::ex_set_2nd_rrfdif_clk_en timed out when waiting for the RRF chassis to respond to a request. This may happen occasionally during a scan. Device Mask = 0x1 If this occurs often you may have RRF chassis communication problem.->

2247160 RRF Fiber Cable Link Not Ready See Long Description Fri Jan 25 09:23:14 2008 Error: 2247160 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 272 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51176 Object: APS Test Task RRF FIBER CABLE LINK NOT READY The fiber link between the RRF and MGD chassis is not in a ready state after hardware initializ

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2292267-2 5V RF MODULE RF POWER SUPPLY 2292267-2 5V RF MODULE RF POWER SUPPLY 2292265 5V RF MODULE DIGITAL POWER SUPPLY
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.