Pular para o conteúdo principal
GE Healthcare · MROptima22674312267446

TPS Reset fails with NO XGD MODULE FOUND Error Due to 48 V Gradient Control Breaker at the PDU Was Tripped.

Symptom

Cannot reset TPS with IRF3 not ready message Cannot reset TPS with IRF3 not ready message.

Error message

2267431 The IRF3 is not ready See Long Description Error: 2267431 Host: MR02 Process: NSPFile: AGP:Irf3RrfHWO Line: 1209Exception Class: Unknown Severity: UnknownFunction: Software/Firmware/Config: N/APSD: Unknown Coil: Unknown Scan: -1/-1 The IRF3 is not ready.-> Extended Error: This is just a warning message that indicates that another error message has already been logged about a serious problem with the IRF3, DTX, RRX, TDM, interconnect cable, or other error that is preventing the system from controlling the exciters and receivers. Look for a previous error message about configuration errors, cable errors, module errors, or calibration errors. Scans will not be allowed until that problem is corrected

2267446 The XGD module did not send a Fiber Link Init Complete

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues