Scanner Hardware Reset Failed.
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 244553 The STC controller has failed to reset See long description for full error Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset . Error: 244518 Scan hardware reset failed See Long Description for complete error Sun Sep 26 08:30:10 2004 Error: 244518 ctge04 fwmgr HwManager.c 1536 Scanning hardware reset failed .
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2104227 SBX-ETHERNET daughter board - Used on CTi ETC and STC Assy. See long description for details and picture 2104227 SBX-ETHERNET daughter board mounted on the ETC board in the table. Part #: 2104227 Part Type: GP REPAIRABLES Part Weight: 0.91 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: SBX-ETHERNET FRU: Yes Status: Orderable Comments: FLS - FE & HQ: YOU CAN SHIP T3202BE.THIS OLDER REV BD WILL WORK INALL SYSTEMS. THE NEW REV T3202FY WILL ONLY WORK IN NEW SITES WHICH ORIGINALLY HAD T3202FY **UNTIL** AFTER ZD MANDATORY FMI SHIPS LATE IN 1995. MB 6/95 4/18/01- T3202FY PLANNED FOR DELIST. SOURCING WAITING QUOTES FOR PRICING/LEADTIME. NO POOL ADDS. MS12464 REPLACED BY:T3202FY GELS - FOR SOFT 1.3 OR + ONLY; 46-264832P1 CAN NEVER REPLACE 2104227 2104227 CAN REPLACE P1 IF 1.3 INSTALLED. SP/MB,PCB SCA-LAN
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required