Application Software does not complete the boot process with Failed to determine hardware state for AGP Due to AC Distri…

Application Software does not complete the boot process with Failed to determine hardware state for AGP Due to AC Distribution ASM.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2218592 The MrMail send of real time filter to TPS timed out Thu Jun 3 05:32:17 2004 2218592 qstmr1 scn msaUtil.m 175 The MrMail send of real time filter to TPS timed out.

2218643 Dummy packet was sent to TPS by Scan because Scan couldn't get Start Recon reply packet. Wed Dec 17 08:24:16 2008 Error: 2218643 Host: MR3T Process: scn File: MsaScanMgr.m Line: 1686 Dummy packet was sent to TPS by Scan because Scan couldn't get Start Recon reply packet.

2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung.

2223625 Send packet opcode:10168 seqnum:7 rev:2 type:1 length:0 req:WavRec resp:SPU status :0 failed. Mon Nov 22 10:57:41 2004 Error: 2223625 Host: MRI1-OC0 Process: WavRec File: mrmailmgr.c Line: 791 Send packet opcode:10168 seqnum:7 rev:2 type:1 length:0 req:WavRec resp:SPU status :0 failed.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
5111201 AC distribution asm excite3 120v 5111201 AC distribution asm excite3 120v
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required