RRF Calibration Error Due to a Bad Connection From RRF MUX to RRF Exciter.
Symptom
Will not scan with error 2248217 Receiver calibration error Fri Jul 18 05:59:36 2003 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APSReceiver calibration error(s). Receiver DB out of spec.R1 values that failed:Rx1: 12 10Rx2: 12 11 10 9Rx3: 12 11 10 9Rx4: 12 11 10Rx5: 12 11 10 9Rx6: 12 11 10 9Rx7: 12 11 10 9Rx8: 12 10
Error message
2248259 VERY LOW SIGNAL STRENGTH IN RRF CAL. Low or no Signal Strength Seen From Exciter 0 (0=Master,1=Slave) Mon May 3 13:52:03 2004 Error: 2248259 Host: lx-mr Process: NSP File: APS:RrfCalibrator::excCal Line: 2541 VERY LOW SIGNAL STRENGTH IN RRF CAL Low or no signal strength seen from Exciter 0 (0=Master,1=Slave) during RRF cal. 2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2297729-3 Cable RRF Slot 7R-J6 RRF Master Exciter J6 to Mux Board J10
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.