Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT182407

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Table will not move up or down during a elevation table reference. Table will not move up or down during a elevation table reference

Error message

Error: 182407 Velocity error detected when trying to find the Index Pulse Sun Aug 29 13:25:59 2004 Error: 182407 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 174 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 13:25:59:690 Log Series: 34 HW/SW error: Velocity error detected when trying to find the Index Pulse. Hardware Inhibit Switches: 0xc0d8 Elev Direction: 0 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2113411 Table Power Supply 170 vdc used for elevation / tilt amp operation see long description for details 2113411 Table Power Supply 170 vdc used for elevation / tilt amp operation Part #: 2113411 Part Type: GP CONSUMABLES Part Weight: -61819.38 LBR Repairable: No Returnable: No Hazardous Material: N Description: 170 VOLT POWER SUPPLY FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues