GE Healthcare · CTBrightSpeed260007323
BrightSpeed HPower, Delight, Velocity or MERI HPower
Symptom
Scanning aborted with DCB errors
Error message
Error: 260007323 Unacceptable Amount of Acquisition Trigger Jitter See LD for complete error Fri May 14 07:25:51 2004 Error: 260007323 Host: DAS Process: DCB Firmware File: abort.c Line: 315 Function: DATA ACQUISITION : Scan Data Trigger Generation Scan: 4371/1/1 Type: Axial Exception Level: Sec/Soft Time: 07:25:51:075 Log Series: 51 Scan ABORT -- DAS Has Detected An Error.^B Acquisition hardware (DCB) has detected an Unacceptable Amount of Acquisition Trigger Jitter. (Trigger Frequency: 984 hertz). (State: Normal State Command Level: Scan in Progress ScanState: Collecting Offsets State)
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2361853-2 GDAS DCB 16-SLICE replaces 2361853 See LD for more Info 2361853-2 GDAS DCB 16-SLICE replaces 2361853 Part #: 2361853-2 Part Type: GP CONSUMABLES Part Weight: -59999.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS DCB FRU: Yes Status: Orderable FLS - GDAS DCB (16-SLICE) USED ON: LIGHTSPEED PRO 16 GDAS ADDED PER GDAS RSPL MS13452 This board has been replaced by 2361853-2, FE NOTE: Site MUST have M4FMI completed before this board will function properly, ms13452
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required